User`s manual
REPEATABILITY TEST
Before starting the straightness measurement it is
recommendable to perform a repeatability test. See the
chapter Repeatability Test.
Do the Repeatability Test at the position far from the laser transmitter.
MEASUREMENT POINT REGISTRATION
1. Position the detector unit as close to the laser transmitter as possible and
open the shutter. Note the angle value displayed on the screen.
Register the measurement value by touching the register icon.
The first measurement point is now registered and the value is
automatically set to 0 (reference point)
You can zero the detector value, and half it, by touching the
corresponding icons before you register the value. If you choose not
to use this option the icons will not appear in the rest of the
procedure.
82