User`s manual

CALIBRATION OF THE SPIRIT LEVELS
Position the T210 on a table with flat surface which is in level within 0,2
mm/m in both directions. Mark two positions for the receiver at a distance of
1 metre minimum from each other.
Min 1 metre between the detector positions.
Zero the levels with the micrometre screws.
Zero the value on the screen.
Read and note the displayed value.
Turn the T210 180°.
Turn the turret as shown.
Zero the levels with the micrometer screws.
Zero the value on the screen.
Read and note the displayed value.
233