User`s manual

EXPORT FUNCTION
With the Export function it is possible to transfer stored measurements to a
computer. The description below covers how to export using the program
Hyper terminal in Windows.
SET-UP HYPER TERMINAL
Start new connection according to your Windows manual.
Enter name, example Fixturlaser, and choose any icon.
Connection
Connect using
: COM 1 (the port where cable is connected).
Port settings
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None
Select file, properties.
Emulation: ANSI
Backscroll buffer lines: 500 (or optional)
Select file and save.
TRANSFER PROCEDURE
Connect the display unit to the COM-port on a computer with a serial null-
modem cable.
Open the Hyper terminal in the computer, example Fixturlaser.ht.
Select transfer, capture text and start.
Touch the Export icon in the Memory manager in the display unit. Wait until
the stored measurements are transferred.
Select transfer, capture text and stop, followed by disconnect under call.
215