Alignment of measurement point
When alignment of measurement point is selected, the measurement point
alignment screen is shown. The actual values for the selected point go live
and alignment can be made against zero. Zero will be in accordance to
selected references.
Note: It is recommended to set the screen filter to 0 when aligning.
1. Position the receiver at the measurement point that should be aligned.
Scroll to the corresponding measurement point in the detailed screen.
Go to the measurement point alignment screen by touching the
alignment icon in the detailed screen.
’
Measurement point alignment screen
2. Align the selected measurement point to zero. Touch OK to confirm
adjusted values.
The program returns to the detailed diagram with the adjusted values.
Note: Depending on your application, alignment at one point might affect
other measurement points. It is therefore recommended to remeasure all
points when all adjustments are made.