Service manual

Ultraview 1030/1050 Monitors — Service Manual
5-2
System Startup
When the unit is first powered ON, the following three window system startup screen is displayed.
Figure 5-1: System Startup window
When this screen is displayed, the unit automatically begins power ON diagnostics. Any
diagnostics failures will be reported in the upper left-hand window. Provided no serious diagnostics
failures are encountered, the unit will continue past this screen and start normal monitor operation.
A “3-2-1 countdown” is also shown in the System Startup window, during which time the booting
operation can be halted and a Boot menu of boot options can be entered.
The Boot menu, and how to enter it, and the power ON diagnostics are explained in the following
sections.
This window displays basic system configuration
This window displays any diagnostic
failures and other system errors
during the boot process.
This window is reserved for future use.
and booting status during the boot process.
The Boot menu and Diagnostics menus are also shown here.
3-2-1 countdown
appears here