Ultraview 1030/1050 Monitors — Service Manual
4-23
Maintenance
Removing the EMI Interface PCBA for Option H
Figure 4-17: PCBA support plate and EMI interface PCBA for option H
1 Remove the four 6-32 × 1/4 inch screws holding the PCBA support plate to access the EMI
interface PCBA assembly.
2 Remove the label directly over the assembly on the outside of the case.
3 Remove the four flathead screws under the label.
4 Unplug the cable from J201 of sensor interface PCBA taking appropriate precautions and
orientation.