Service manual

Ultraview 1030/1050 Monitors — Service Manual
4-20
Capnograph with Ethernet Removal/Disassembly — Option
G or H
Removing the I/O Bay PCBA
Figure 4-12: Ethernet removal and disassembly
1 Remove the four screws mounting the case to the 90367/90369 housing (two each
6-32 × 1.125 inch; one each 6-32 × 1/2 inch; and one each 10-32 × 1/2 inch).
2 Remove the case and retain the gasket for re-assembly.
Figure 4-13: Jackscrews and knurled nuts