Service manual

Ultraview 1030/1050 Monitors — Service Manual
4-14
Replacing the Fan/Battery Contact Assembly
1 Remove the chassis assembly from the housing as previously described.
2 Separate the chassis assembly from the module wall assembly by removing the two screws
above the battery ejection springs and the two screws near the bottom.
Figure 4-6: Replacing the fan/battery contact assembly
3 Pry up the center retainer on the battery contact assembly (near the battery ejection springs).
This should free the fan/battery contact assembly.
4 Remove the screw between the chassis and the interconnect PCBA.