Ultraview 1030/1050 Monitors — Service Manual
4-2
Remove excess liquid from the cloth and squeeze dry.
Wipe exposed surfaces of the equipment and cables.
Remove any soap residue by gently wiping with a clean damp cloth.
Wipe dry with a clean dry cloth.
Use only the following recommended cleaning solutions:
Mild soap and water solution
U.S. Pharmacopoeia (USP) green soap
Sodium hypochlorite solution (1:10 dilution of household chlorine bleach in water)
Phenolic germicidal detergent (1% aqueous solution)
Glutaraldehyde (2.4%) (Cidex)
Isopropyl alcohol (70% solution)
Note:
Over time, repeated use of a chlorine bleach solution may cause some colors to fade.
Tape adhesive can be removed with Spacelabs Medical adhesive tape remover pads
(P/N 392196-001).
Questions and concerns about cleaning issues should be directed to a Spacelabs Medical field
service engineer.
Fan Filter
Check that the fan filter is clean. If required, clean the filter by washing it with mild soap and water.
Monitor, Expansion Housing, Module Housing, and Recorder
Clean the case by washing it with mild soap and water or use Plast-N-Glas cleaner. Use TF
solvent for cleaning the electronic connectors and contacts as necessary.
Note:
Avoid directly spraying liquids into the recorder openings.