Service manual

Contents
4-1
Maintenance
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Cleaning/Disinfecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Mechanical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Electrical Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Capnography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Part Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Overview
Caution:
Observe precautions for handling electrostatic-sensitive devices!
Note:
Never touch electrostatic-sensitive electronic components without following proper
anti-static procedures, including the use of an ESD wrist band and mat. An electrostatic
discharge from your fingers can permanently damage electronic components.
All static-sensitive electronic components are packaged in static-shielding bags. Retain the
bag for repackaging the component, should you need to store it or return it to Spacelabs
Medical for any reason.
Cleaning, preventive maintenance, and safety checks should be performed annually and
following any product disassembly/assembly. Preventive maintenance and safety checks
must be performed by trained personnel only.
Cleaning/Disinfecting
Warning:
Use only recommended cleaning solutions, or you may void the manufacturer’s
warranty.
Harsh chemical agents degrade plastics and will compromise the safety of the
device.
Disconnect the equipment from the patient and the electrical supply before cleaning.
Do not allow liquid to enter the interior of the module or monitoring equipment.
Do not immerse the equipment or cables in water or cleaning solutions.
Do not autoclave.
To clean the exterior of monitors, modules, and cables:
Prepare the cleaning solution according to the manufacturer’s instructions.
Wet a clean cloth with the selected cleaning solution.