Battery Chargers
The battery chargers have the following features:
Front panel LED battery charging/failure indicator.
Automatic “battery care” recharge cycles to ensure a maximum charge.
One or two batteries fully charge in 90 minutes if the system is not operating; 4 to 5 hours
while unit is operating.
Individual battery chargers for each of the 2 batteries so that a bad battery will not degrade the
other battery. Will not prematurely “kill” batteries.
Figure 3-7: Charging circuit block diagram
12-to-24 V
DC input
charger pre-regulator
charger power converter
charger CPU IC
DC output
20 V
DC to DC converter
3.6 A max current at
20 VDC. MIC38HC43
IC chip, 100 KHz
Compute battery condition
Turn ON/OFF power converter IC
CPU controlled battery
battery charge rate
1.8 A max, 0.16 A min
Analog feedback
Using DC charge levels and
an AC waveforms, evaluate
battery internal condition
+
charger output to battery
+
5
v
+5 V
BTI CCS9620 IC chip
Convert 20 V to 10 V min and
15.1 V max; constant current
limited with constant voltage
limited; LT1511 IC chip
12 V lead
acid battery
battery fault