Service manual

Ultraview 1030/1050 Monitors — Service Manual
2-31
Setup
Configuration Options for Wireless Ethernet Option (90310)
Figure 2-8: 90310 Wireless Ethernet Interface PCBA
Master/Slave
Selects if 90310 is configured as a master or slave monitor. Only one master unit is allowed
per channel selection.
A/B
Unused
90309/90310
Defines the unit where the wireless PCBA is installed.
Normal/Loop
Configures the diagnostics mode. In “normal”, only one diagnostics pass is completed. In
“loop”, diagnostics will loop continuously after displaying error conditions.
Thumb Wheel Switch
Configures the network channel number. There are 16 positions, 0 through F. 0 allows the unit to
roam to the first available master. 1 through F identify a specific network channel.
configuration jumpers (see below)
thumb wheel switch