Service manual

Ultraview 1030/1050 Monitors — Service Manual
2-29
Setup
Verify that the ECG waveform is displayed.
3 Connect a waveform simulator to the pressure inputs.
4 Label and zero the channels.
Verify that the numerics and waveforms are accurate.
5 Touch RECORD, then touch one of the flashing parameter keys.
Verify that the recording is printed.
Note:
If the monitor is installed on a network, perform this test as described. Verify that the
recordings are printed at the central printer.
Optional 90310 Wireless Network Interface (WNI)
90310 Front Panel
Rack mounting hardware — mounts the 90310 housing onto the housing pedestal mounts
Single housing kit (P/N 016-0365-00)
Double housing kit (P/N 016-0364-00)
Power LED — when lit, indicates the presence of DC voltage to the 90310