Ultraview 1030/1050 Monitors — Service Manual
2-22
FLASHING/STEADY ON Key
The FLASHING/STEADY ON key selects the mode of operation for the alarm relay ( ). In
FLASHING mode, the alarm relay contacts open and close according to the priority of the active
alarm level. In the STEADY ON mode, the relay contacts close if any alarms are active. They open
only when all alarm conditions cease.
ALARM LEVEL Key
The ALARM LEVEL key defines the level of alarms responded to by the monitor (activate alarm
relay): high, medium, or low. Touching ALARM LEVEL displays the Alarm Level menu:
Select HIGH to monitor only high priority alarms.
Select MEDIUM to monitor medium and high priority alarms.
Select LOW to monitor all alarm priorities (high, medium, and low).
MINIMUM VOLUME ON/OFF Key
The MINIMUM VOLUME ON/OFF key locks (ON) the current alarm tone volumes as the new
minimum volumes or unlocks (OFF) the tone volume controls.
QRS/SPO2 TONE ENABLE Key
The QRS/SPO2 TONE ENABLE key defines whether the QRS or SpO
2
tones, when enabled
within the ECG or SPO2 menus, sound all the time (Always) or only when this monitor is in alarm
(During Alarm).
ALARM WATCH TONE Key
The ALARM WATCH TONE key defines whether the alarm watch notification tone sounds once or
three times whenever this monitor receives a new alarm watch multicast message.
UNITS OF MEASURE Key
Touch the UNITS OF MEASURE key to select a unit of measure for the monitor:
The highlighted option is the one that is currently active. Touching another key changes the
selection.
HIGH MEDIUM
LOW
ALARM LEVEL - select minimum alarm priority level to trigger relay activation
inches
cm
Restart monitor after selecting units of measurement
mmHg
kPa
lb
Kg