Service manual
Ultraview 1030/1050 Monitors — Service Manual
2-21
Setup
THIS MONITOR includes all the parameters on this monitor.
1 Touch the appropriate option.
2 Return to step 1 to define the remaining pre-selected recordings.
ALARM SETUP Key
The ALARM SETUP key enables you to configure the alarm features. Touching this key displays
the Alarm Setup menu:
REMOTE ACCESS ON/OFF Key
The REMOTE ACCESS ON/OFF key enables (ON) or disables (OFF) a remote monitor’s access
to this monitor’s alarm limits.
ALARM SUSPEND ON/OFF Key
The ALARM SUSPEND ON/OFF key enables (ON) or disables (OFF) operation of the monitor’s
alarm suspend function. With the key ON, ALARM SUSPEND will be displayed on the key. With
the key OFF, TONE RESET will be displayed on the key.
TREND SUSPEND ON/OFF Key
When the TREND SUSPEND ON/OFF key is set to ON, trending is disabled when alarms are
suspended using the Alarm Suspend hard key. When this key is set to OFF, trending operation
becomes independent of the Alarm Suspend hard key (i.e., the key will always trend when a
parameter is active).
ALARM RELAY Key
Touching ALARM RELAY displays the Alarm Relay Setup menu:
RELAY TIME-OUT 0 SEC/10 SEC Key
The RELAY TIME-OUT key sets the number of seconds (0 or 10) that the alarm relay remains
closed following the end of an alarm condition.
ALARM SETUP - Select parameter to change
REMOTE ACCESS
ON OFF
TREND SUSPEND
ON OFF
ALARM
RELAY
QRS/SPO2 TONE ENABLE
ALWAYS DURING ALARM
ALARM WATCH TONE
MORE
SINGLE TRIPLE
ALARM SUSPEND
ON OFF
MIN VOLUME
ON OFF
FLASHING
ALARM
LEVEL
RELAY TIMEOUT
0 SEC 10 SEC
ALARM RELAY SETUP - Select parameter
STEADY ON