Service manual
Ultraview 1030/1050 Monitors — Service Manual
5-24
26 EEPROM Tests - New checksum no good RCDIAG
38
Printer Tests - WRRDY inactive or -SYNC active
after reset
RCDIAG
39
Printer Tests - Readback system wrong state after
reset
RCDIAG
3A Printer Tests - Error active after reset RCDIAG
3B Printer Tests - +BUSY not active after idle CMD RCDIAG
3C Printer Tests - +BUSY not inactive after idle CMD RCDIAG
3D Printer Tests - Error not active after illegal CMD RCDIAG
3E Printer Tests - Error not inactive after idle CMD RCDIAG
3F Printer Tests - +BUSY not inactive after idle CMD RCDIAG
40 Printer Tests - Incorrect feedback after illegal CMD RCDIAG
41 Thermal array over-temperature RCDIAG
42 Low 12 V supply to AR42 recorder RCDIAG
43 High 12 V supply RCDIAG
44 Other AR42 error RCDIAG
27 Error detected during print test RCDIAG
2A SDLC cable not present RCDIAG
2B SDLC clock frequency too high RCDIAG
2C SDLC clock frequency too slow RCDIAG
2E Watchdog timeout too short RCDIAG
2F Watchdog timeout too long RCDIAG
D1 Stack error RCMAIN
67
Printer error code - Printer did not like command
sent
RCKEYP
Table 2: Recorder Error Codes (continued)
Error
Code
(hex)
Failed Assembly or Likely Cause Program Modules