Ultraview 1030/1050 Monitors — Service Manual
5-22
During normal operation, D2 will be dimly lit (it is actually flashing very rapidly). LED Error Codes
are shown in Table 2.
Table 2: Recorder Error Codes
Error
Code
(hex)
Failed Assembly or Likely Cause Program Modules
80 EPROM checksum bad
Recorder Code
Diagnostics
(RCDIAG)
8F Stack pointer bad RCDIAG
81 Internal RAM won’t write all zeroes RCDIAG
82 Internal RAM won’t write all ones RCDIAG
83 Timer 0 - No overflow RCDIAG
84 Timer 0 - High byte bad RCDIAG
85 Timer 0 - Low byte bad RCDIAG
86 Timer 1 - No overflow RCDIAG
87 Timer 1 - High byte bad RCDIAG
88 Timer 1 - Low byte bad RCDIAG
30 SDLC Errors - Bad Xmit status RCDIAG
31 SDLC Errors - No SI RCDIAG
32 SDLC Errors - No start flag RCDIAG
33 SDLC Errors - Bad address RCDIAG
34 SDLC Errors - Bad frame type RCDIAG
35 SDLC Errors - No end flag RCDIAG
10 Stuck I/O bus bit 0 low RCDIAG
11 Stuck I/O bus bit 1 low RCDIAG
12 Stuck I/O bus bit 2 low RCDIAG
13 Stuck I/O bus bit 3 low RCDIAG