Service manual
Ultraview 1030/1050 Monitors — Service Manual
5-14
01030B06 IRTS uP RAM failed Replace CPU PCBA.
01030B07 IRTS uP ROM failed Replace CPU PCBA.
01030B08 IRTS data dump failure Replace CPU PCBA.
01030B09 Interactive IRTS test failure Replace CPU PCBA.
01030B0A IRTS ALU error Replace CPU PCBA.
01030B0B IRTS Remote Keypad ALU error Replace CPU PCBA.
01030B0C
IRTS ALU warning status, please
calibrate
Re-calibrate the touchscreen
and retest. If problem
persists, replace IRTS PCBA/
bezel.
01030C00 OS error while diagnosing KBD
Reboot and retest. If problem
persists, replace CPU PCBA.
01030C01 Can’t access KBD device registers Replace CPU PCBA.
01030C02 KBD didn’t respond to command Replace CPU PCBA.
01030C03 KBD failed selftest Replace CPU PCBA.
01030C04 KBD failed interface test Replace CPU PCBA.
01030D00 Cant access AUDIO device registers Replace CPU PCBA.
01030D01 Could not open audio device
Reboot and retest. If problem
persists, replace CPU PCBA.
01030D02 Audio device didn’t respond correctly Replace CPU PCBA.
01030D03 Audio Loopback test failed Replace CPU PCBA.
01030E00 NVRAM Test Failed Replace CPU PCBA.
01030E01 NVRAM Clock Not Running Replace CPU PCBA.
01030E02 NVRAM Checksum error
Zero the NVRAM and reboot.
If problem persists, replace
CPU PCBA.
Table 1: Diagnostic Failure Messages (continued)
Error Code Diagnostics Failure Message Suggested Action