Service manual
Ultraview 1030/1050 Monitors — Service Manual
5-13
Troubleshooting
01030700 PCI Bridge Test Failed Replace CPU PCBA.
01030701
PCI Bridge configuration not
recognized
Replace CPU PCBA.
01030702 PCI Bridge registers are not writable Replace CPU PCBA.
01030703
PCI Bridge revision not valid for
clinical use
Replace CPU PCBA.
01030800 Ethernet Test Failed Replace CPU PCBA.
01030801 Ethernet Setup Failed Replace CPU PCBA.
01030802 Ethernet Transmit Failed Replace CPU PCBA.
01030900 ISA Bridge Test Failed Replace CPU PCBA.
01030A00 Video Test Failed Replace CPU PCBA.
01030A01 Invalid Display Type Replace CPU PCBA.
01030A02 Invalid Display Size Replace CPU PCBA.
01030A03 Video configuration not recognized Replace CPU PCBA.
01030A04 Video DRAM failure Replace CPU PCBA.
01030A05 Video could not detect hsync signal Replace CPU PCBA.
01030A06
Video could not detect blue video
signal
Replace CPU PCBA.
01030B00 Couldn’t open IRTS device driver
Reboot and retest. If problem
persists, replace CPU PCBA.
01030B01 IRTS interface failed Replace CPU PCBA.
01030B02 IRTS ring PCBA interface failed Replace CPU PCBA.
01030B03 IRTS IR pair failed
Re-calibrate the touchscreen
and retest. If problem
persists, replace IRTS PCBA/
bezel.
01030B04 IRTS reset failed Replace CPU PCBA.
01030B05 IRTS change mode failed Replace CPU PCBA.
Table 1: Diagnostic Failure Messages (continued)
Error Code Diagnostics Failure Message Suggested Action