8 - Troubleshooting
A) Electrode error
The Compex 3 will beep and alternately
display a pair of electrodes and an arrow
pointing to the channel where the problem
was detected. In the example above, the
stimulator has detected an error in
channel 1.
The electrode error message can mean
that:
- there are no electrodes connected to this
channell,
- the electrodes are old, worn, and/or the
contact is poor: try using new electrodes,
- the electrode cable is faulty: try
connecting it to another channel. If the
problem persists, replace the cable.
If this message appears, disconnect the
charger. The unit will automatically turn off.
Contact your customer support service
and quote the corresponding error code.
If you experience any other problems with
your unit, contact your nearest authorized
customer service support service.
B) Memory full
Programme statistics are stored in the
Compex 3’s memory as compiled. When
this memory is full, the following screen will
appear.
38a: Press the On/Off button to turn off
the unit.
38b: Press the channel 1 +/- button to
access the screen shown in Figure 39 and
empty the memory without saving the
data.
38e: Press the channel 4 +/- button to
continue using the stimulator without
emptying the memory. If you choose this
option, no further programme data will be
stored.
NOTE: If you are using Compex 3
software, you can store all the data from
your stimulator in your PC. (See
"Sinchronizing" in the "Compex 3 Software"
chapter.
39a: Press the On/Off button to turn off
the unit.
39b: Press the channel 1 +/- button to
return to the previous screen.
39e: Press the channel 2 +/- button to
confirm your choice. This will erase all the
data stored in the Compex 3 without
storing it in your computer.