User manual
29
COMPEX 3 User Manual
24a: Press the On/Off button to return to
the previous screen
24e: Use the channel 4 +/- button to confi
rm your choices and launch the
programme.
NOTE: Current intensity is automatically
calculated in line with the size of electrode
used. Electrode size is determined by
colour. Make sure you choose the correct
colour before running the programme.
Press the i button from the stimulation
screen or the end of programme screen to
access the programme parameters.
B) Hyperhidrosis
Read the chapter “Hyperhidrosis”
in the Practical Guide to become fully
familiar with this programme before using
it.
NOTE: The Hyperhidrosis programme
should always be run in channel 1 with the
large (red) iontophoresis electrodes
supplied with the stimulator.
Use the 2.1 mm pin-connector cables
(grey) to connect the electrodes
The Hyperhidrosis programme is in the
Direct Current category of the Specific
Treatment type.
25a: Press the On/Off button to return to
the previous screen.
25b: Use the channel 1 +/- button to
choose a programme.
25e: Press the channel 4 +/- button to
confirm your choice and execute the
programme.
26a: Press the On/Off button to return to
the previous screen.
26c: Use the channel 2 +/- button to
change the default electrical density value.
26e: Press the channel 4 +/- button to
confirm your choices and access a
confirmation screen.
Use the i button to access the electrode
placement pictogram.
The
Hyperhidrosis programme lasts for
12 minutes and can only be run on channel
1. Only use the red electrodes with this
programme. Other electrodes will cause
current intensity calculation errors.
27a: Press the On/Off button to return to
the previous screen.
27e: Press the channel 4 +/- button to
confirm your choices and launch the
programme.
Use the i button to access the electrode
placement pictogram.
NOTE: Press the i button from the
stimulation screen or the end of
programme screen to access the
programme parameters.
C) Oedema
Read the chapter “Oedema” in
the Practical Guide to become fully familiar
with this programme before using it.
The Oedema programme is in the Direct
Current category of the Specific
Treatment type.
f
ig. 24
fig. 25
f
ig. 26
fig. 27
24a 24e
25a 25b 25e
26a 26c 26e
27a 27e