User manual
24
COMPEX 3 User Manual
m-4 connected
The stimulator will automatically select the
muscle group to be stimulated when a
m-4 cable is connected to the unit.
Reminder: The m-3 function will activate
automatically in programmes that
accommodate this function when at least
1 m-4 cable is connected to the
stimulator.
8b: Use the channel 1 +/- button to enable
the m-1 function. When the function is
enabled, the screen will display the word
ON rather than OFF. When in ON mode, it
is the patient who triggers the stimulation.
Reminder: Not all motor nerve stimulation
programmes use the m-1 function. If
the m-1 function is not available, there
will be no symbol above channel 1.
7- 8c: Press the channel 2 +/- button to
enable the TENS function. When enabled,
channels 1 and 2 are used for the
programme chosen, and channels 3 and 4
for the TENS treatment.
NOTE: When the TENS function is enabled,
the m-1 cannot be activated and the
corresponding symbol will disappear
automatically from the screen.
Use the channel 2 - button to delete the
default warm-up period.
7 - 8d: Use the channel 3 +/- button to
choose the programme difficulty level.
7 - 8a: Press the On/Off button to return
to the previous screen.
7 - 8e: Use the channel 4 +/- button to
confirm your choices and launch the
programme.
E) During the stimulation session
Preliminary m-3 test
This test is launched only if a cable fitted
with the m-4 system is already
connected to the stimulator when this is
turned on!
It is essential that the patient
remains completely still and relaxed during
the entire test.
If the m-4 cable is connected, the test
will begin as soon as you choose and
personalise your programme.
NOTE: The test is not launched if you
choose a programme that does not
require the preselection of a muscle group.
9e: The m logo above channel 4 indicates
that the m-4 cable is connected to the
stimulator. The figures in channel 4 will
automatically change during the test as
the sensor records different
measurements.
NOTE: When several sensors are
connected simultaneously, the m-3 test
is performed progressively, one m-4
connected channel after the other. The
channel being scanned is easily identified
by its changing numbers and the mi logo
above the channel.
The 4 +/- buttons will be disabled for the
duration of the test.
9a: Press the On/Off button to interrupt
the test at any moment. To relaunch the
test from the start, press any of the 4 +/-
buttons.
Press the On/Off button a second time to
return to the list of programmes.
Once the test is complete, you will
automatically be taken to the default
programme start screen and prompted to
increase the stimulation energies.
S
ession duration
(in min.)
TENS programme enabled
Combination without TENS programme
fig. 8
fig. 9
no warm-up sequence
warm-up sequence
8a 8b 8c 8d 8e
9a 9e