User's Manual

17/21
IEC61000-4-8
a typical location in a
typical commercial or
hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test voltage.
Table 3: Electromagnetic immunity 2
Guidance and manufacturer's statement - Electromagnetic immunity
IR-FT Infrared Forehead Thermometer is intended for use in the electromagnetic environment as
specified below. The buyer or user of IR-FT Infrared Forehead Thermometer should assure that it is
used in such an environment:
Immunity test
IEC 60601 Test level
Compliance
level
Electromagnetic environment -
Guidance
RF conduction
IEC61000-4-6
RF radiation
IEC61000-4-3
3 V (r.m.s.)
150 kHz~80 MHz
3 V/m
80 MHz~2.5 GHz
Not applicable
3 V/m
Portable and mobile RF communication
equipment shall be used not closer to
any part of IR-FT Infrared Forehead
Thermometer, including cables, than the
recommended separation distance. The
distance shall be calculated according to
the formula applicable to the frequency
of the transmitter.
Recommended separation distance
d =
P2.1
d =
P2.1
80 MHz~800 MHz
d =
P3.2
800 MHz~2.5 GHz
where:
P
The maximum rated output power
of the transmitter in watts (W) according
to the transmitter's manufacturer;
dThe recommended separation
distance in m
b
.
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
c
, should be
less than the compliance level in each
frequency range
d
.
Interference may occur in the vicinity of
equipment marked with the following
symbols.
Note 1: At 80 MHz and 800 MHz, the formula used for higher frequency range applies.
Note 2: These guidances may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and human bodies.