12. Clinical accuracy:
The normal body temperatures are within a certain range. The body temperatures measured at
different sites will be different. See the specific differences in the following table:
Measurement site
Normal body temperature
Deviation
Forehead
temperature
34.7℃~37.8℃
(94.5℉-100℉)
0.3℃~0.5℃ (0.5-0.9℉)
lower than the oral
temperature
The product is applicable to people of all ages. The body temperatures of different groups will be
different. See the specific differences in the following table:
Age group
Normal body temperature
Clinical deviation
At the age of 0-2
36.4℃~38.0℃
(97.5-99.5℉)
≤±0.3℃(0.5℉)
At the age of 3-10
36.1℃~37.8℃
(97-100℉)
≤±0.2℃(0.4℉)
At the age of 11-65
35.9℃~37.5℃
(96.6-99.5℉)
≤±0.2℃(0.4℉)
At the age of 65 above
35.8℃~37.4℃
(96.4-99.2℉)
≤±0.3℃(0.5℉)
Caution: Depending on the person's skin color, thickness and body part and great changes of
ambient temperature, the measured body temperature valueswill be different, which is quite normal.
The reason is that the body parts more frequently exposed to the external environment will be more
affected by the ambient temperature.
13. Statement
The product is intended to measure heat radiation from the forehead to display body temperature. It
has been verified for clinical accuracy. The performance verification method for the product can be
accessed by contacting our Customer Service Department.
14. Description on the replacement cycle and method of consumables and batteries
The detachable components of this product mainly include the batteries. Before installing the
accessories, please confirm whether the specifications of accessories are consistent with the original
specification. Please use the components of models specified by the company. Contact the dealer or
manufacturer if necessary.
Battery mounting and replacement: