See details of guidance and manufacturer's statement in the subsequent “Guidance and
Manufacturer's Statement”.
Warnings:
IR-EFT ThermArt shall not be used adjacent to or stacked with other equipment, and if adjacent
or stacked use is necessary, it shall be verified the device can work in normal normal
operation in the configuration in which it will be used;
If the physiological parameters measured by the equipment are lower than the specified
minimum values, the measurement results may be inaccurate.
Table 1: Electromagnetic emission
Guidance and manufacturer's statement - Electromagnetic emission
IR-EFT ThermArt is intended for use in the electromagnetic environment as specified below.
The buyer or user of IR-EFT ThermArtshould assure that it is used in such an environment:
Emission test
Compliance
Electromagnetic environment - Guidance
RF emission
CISPR 11
Group 1
IR-EFT ThermArt uses RF energy only for its internal
function. Therefore, RF emissions are very low and
are not likely to cause any interference to nearby
electronic equipment.
RF emission
CISPR 11
Class B
IR-EFT ThermArtis suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emission
IEC61000-3-2
Not
applicable
Voltage
fluctuation/flicker
emission
IEC61000-3-3
Not
applicable
Table 2: Electromagnetic immunity 1
Guidance and manufacturer's statement - Electromagnetic immunity
IR-EFT ThermArtis intended for use in the electromagnetic environment as specified below. The
buyer or user should assure that it is used in such an environment:
Immunity test
IEC 60601 Test Level
Compliance level
Electromagnetic
environment -
Guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV, contact discharge
±8 kV, air discharge
±6 kV, contact
discharge
±8 kV, air
discharge
Floors shall be wood,
concrete or ceramic
tile. If floors are
covered with the
synthetic material, the