User manual

56-3124-15 Edition AF 2003-08 7
Printing chromatograms
When several chromatograms have been imported into a single window the order in
which they appear on a print out may not be as required e.g. a comparison of several
different runs at different pH may not be displayed in a consecutive numerical order of
pH values.
Action: This is a function of Windows operating system. The problem can be resolved by
importing the curves in the same order as is required for printing and highlighting the
first curve in the series.
Printed method information
When Method Information is printed in a Method Information print, a customized
report or a standard report, the result file name will appear in place of the method name.
Action: Print a report and select the method file name in the Header tab. De-select the
Info item in the Documentation tab.
Evaluation procedure with print
If you run an Evaluation procedure including printing a report, the system
communication may crash and the result file may be saved in the Failed directory.
Action: Do not use automatic evaluation procedures with print.
Curve presentation
Curves displayed as thick, dashed lines on screen are printed as single lines.
Action: No action possible. This is a function of the line drawing routine in Windows
NT.
Landscape printing
In some installations the HP DeskJet 895 will not print documents from UNICORN in
landscape mode.
Action: Do not use the printer driver for HP DeskJet 895 that is supplied with Windows
NT. Select the printer driver for HP DeskJet 855C instead.
UNICORN 3.2 manual, ed.AA, errata
The instruction GoTo_Tube does not exist.
Action: Use the common fractionation instructions.
Resetting the position of the fraction collector for small and large tubes.
Action: Use Reset_Frac_Position.
Using AccumulatorWash instruction for ÄKTAdesign systems (excluding ÄKTAFPLC).
Action: Use manual control to begin flow and turn the Outlet valve to the frac position