User manual
6 56-3124-15 Edition AF 2003-08
Creating methods using block programming
To ensure that the End_block is automatically updated when a gradient length is
programmed or changed the gradient length should be programmed in two steps.
Action:
Select 0 for gradient length, press Insert
Select required gradient length, press Change
Programming Watch functions
The variable parameter Action: Ready does not function when programming Watch
events within the Method Editor.
Action: Use a Watch event that calls for a block containing the instruction Ready.
System Report
If a System report is requested during a run, the program may sometimes crash and a
reboot is required.
Action: Do not request a System report during a run.
Using the instructions Record_on and End_Timer
If you during manual runs execute the instructions Record_on and End_Timer
simultaneously, there is a risk for the run going on infinitely.
Action: Do not execute the instructions Record_on and End_Timer simultaneously.
Blank fields in Instruction box
Occasionally the Instruction box in System Control appears with only blank fields.
Action: Close the Instruction box and reopen it.
Integration
Peak labels and start/end markers shown in the curve window are always calculated from
the latest integration. If an earlier calculated baseline or peak table is selected, the peak
information in the curve window is not updated. Displayed peak labels and markers and
the peak table will not correspond.
Action: Always work with the latest calculated baseline and peak table.
Saving modified result files
When a chromatogram is removed from a result file and the file is then saved the
evaluation module closes.
Action: This is a function of a drive routine in Novell Intranet Ware. This fault has not
been seen when using Novell NetWare v4.50.819 or v4.60.202.
Reports with litre base
If litre is used as the unit, variables will be printed as ml in the variable list when reports
are generated