User manual

56-3124-15 Edition AF 2003-08 3
Software requirements for UNICORN 3.21
Windows NT Workstation 4.0 + Service Pack 5 or later
Network requirements
Supported network cards:
3COM Etherlink III
Compaq™ Netelligent 10/100 TX Embedded UTP Controller
Compaq Integrated NetFlex-3 Controller
AMD PCNET PCI Ethernet Adapter (Integrated)
Novell™ NetWare™ version 4.50.819 or later or Windows NT Server 4.0
A valid network connection
Important advice on operation
Viewing long runs
On long runs that exceed the x-axis limits (21000), the curves window in System Control
may show a false (shortened) x-axis.
Action: After completion of the run, open the result in Evaluation to see the proper curve
window.
Installation
Always re-install Windows NT Service Pack after a UNICORN installation.
Novell Netware problems
The Method Editor may terminate if Novell Netware version 4.83 is used.
Action: Install Service pack 1 for Novell Netware version 4.83.
Logoff / logon Windows NT
UNICORN will not be able to connect to the system(s) if you logoff from Windows NT
and then logon as a different user, followed by starting UNICORN.
Action: Shut down and restart the computer, then start UNICORN and connect to the
system(s).
Restricting the use of the “Continue” function
It is not possible to restrict the ability to use the “CONTINUE” function in the System
Control module. All users are able to use the “CONTINUE” function regardless of their
access rights.
Location of drives
UNICORN cannot locate a drive through the User Setup in Main Menu if the drive is
empty.
Action: Ensure that the drive to be located contains some type of file.
Installation of new strategy
Check default values for System Settings and recalibrate all relevant modules.