User Manual
10 IE-SNC-T03(4)10+T5410-06
Readings on LCD display
°C, °F
Reading next to this symbol is measured temperature or error state of value.
hPa, PSI, inHg,
Reading next to this symbol is measured pressure or error state of value. If selected pressure unit is
mBar or oz/in
2
or mmHg or inH
2
O or kPa, there is shown only value (number) without
corresponding pressure unit!
°C / °F DP
Reading next to this symbol is calculated dew point temperature or error state of value.
symbol 3 near the left edge of the display
Is on if jumper is closed.
Technical parameters of the device:
RS 485 Interface:
Receiver-Input Resistance: 96 kΩ
Devices on bus: max. 256 (1/8 Unit Receiver Load)
Measuring parameters:
Ambient temperature (RTD sensor Pt1000/3850ppm):
Measuring range: -30 to +80 °C
Resolution of display: 0.1 °C
Accuracy: ± 0.4 °C
Atmospheric pressure (T5410 only):
Unit
hPa, mBar
PSI
mmHg
inHg
inH
2
O
oz/in
2
kPa
Range
600
8.70
450.0
17.72
240.9
139.2
60.00
1100
15.95
825.1
32.48
441.6
255.3
110.00
accuracy
T=23 °C
±1.3
±0.02
±1.0
±0.04
±0.5
±0.3
±0.13
0≤T≤40°C
±1.5
±0.02
±1.1
±0.04
±0.6
±0.3
±0.15
Else
±2.0
±0.03
±1.5
±0.06
±0.8
±0.5
±0.20
Response time (air flow approximately 1 m/s):
temperature t90 < 4 min (temperature step 20 °C)
atmospheric pressure
2
: t90 <44s
Recommended calibration interval: T0310, T0410 2 years
T5410 1 year
Measuring interval and LCD display refresh: 0.5 s, atmospheric pressure 2 s
Power: 9 to 30 V dc
2
There is possible to change response time. For more details see file “Description of calibration and adjustment
procedure.pdf”, which is installed together with Users software.