3.  CONNECTION TO PRESSURE SOURCES
The C9500 Series instruments are tted with 
1
/
8
" BSP female connectors. The following checks 
must be made before connecting the instrument to 
a pressure source:-
Check the security of all ttings.
Check that the pressure source does not exceed the 
ratings for the instrument as set out in Table 1.
Check that the gases to be measured will not damage 
or corrode stainless steel, brass, glass, ceramic, silicon, 
nickel, aluminium, gold or epoxy resin.
Note: Comark C9500 Series Pressure Measurement 
Instruments are not suitable for absolute pressure 
measurement.
Note: Pressures in excess of maximum over range 
gure with respect to atmospheric pressure, should 
not be applied to either port as this could damage the 
instrument or represent a safety hazard.
TABLE 1
Instrument  Pressure Range  Maximum Over/ 
    Under Pressure
C9551  0 to ±2 PSI/140mbar  6 PSI/400mbar
C9553  0 to ±5 PSI/350mbar  15 PSI/1000mbar
C9555  0 to ±30 PSI/2000mbar  90 PSI/6200mbar
C9557  0 to ±100 PSI/6900mbar  200 PSI/13700mbar