Operator`s manual

41
LOW PERFUSION
Reusable:
Finger Clip: ±3 digits ±3 digits
Flex: ±3 digits ±3 digits
Soft Sensor: ±3 digits ±3 digits
Reflectance: ±3 digits ±3 digits
Disposable: 6000 Series: ±3 digits ±3 digits
7000 Series: ±3 digits ±3 digits
Alarm Volume:
High:
Low:
75dBA
64dBA
Informational Tone Volume:
High:
Low:
67dBA
55dBA
Measurement Wavelengths and Output Power:
c
Red: 660 nm @ 0.8 mW maximum average power
Infrared: 910 nm @ 1.2 mW maximum average power
Memory
70 hours (assuming continuous operation)
Temperature (Operating):
0º to +40º C (32º F to 104º F)
Temperature (Storage/Transportation): -30º to +50º C (-22º F to 122º F)
Humidity (Operating):
10 to 90% noncondensing
Humidity (Storage/Transportation): 10 to 95% noncondensing
Altitude (Operating):
Up to 12,000 meters (40,000 feet)
Hyperbaric Pressure Up to 4 atmospheres
Power Requirements (Mains)
100-240 VAC 50-60 Hz
Internal Power
Battery: 7.2 volt NiMH battery pack
Operating Life (fully charged
battery):
16 hours minimum
Storage Life: 21 days minimum
Recharge Rate: 4 hours maximum
Dimensions
Approximately 219 mm (8.6”) W x 92 mm (3.6”)
H x 142 mm (5.6”) D
Weight
Approximately 900 grams (2 lbs) with battery
Warranty
3 years
Classification per IEC 60601-1/CSA601.1/UL60601-1 30EM:
Type of Protection:
Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection: IPX2
c.) This information is especially useful for clinicians performing photodynamic therapy.