Playing Back Memory Data
The Model 7500 has a Memory Playback feature, allowing stored data to
be output through an external serial connection. Playing back the data
does not clear the data from memory.
1. With the unit off, connect the serial connector port of the Model 7500
to the back of your computer using the 7500 SC cable, which is
available from NONIN.
2. Press and hold the Plus (+) button while briefly pressing the
ON/STANDBY button.
3. Release the Plus (+) button. Playback mode will be shown on the
SpO
2
and Pulse Rate displays until memory playback is completed.
4. When Memory Playback is complete, the device will return to normal
operation.
NOTES:
Patient memory cannot be cleared when the Model 7500 is in Patient Security
mode.
If using nVISION software, select “Model 2500” for model type (the Model 2500
option also applies to the Model 7500).
E07 is an error code which occurs during memory playback and signifies that
the unit cannot guarantee the accuracy of all patient data in memory. Clear the
device memory to ensure any inaccurate data is deleted. If the error continues
to occur on memory playback, call Nonin Customer Support.