Specifications

A64192AA2 FINAL
8-11
Stockyard Unit
Stockyard Unit Error Recovery
8
4-06 SM06 Failure to move the
Stockyard Transfer Arm
on the second level in
the Y-axis direction.
3000 Tube
Stockyard
To p V i e w
C-4/5 1. Ensure the Stockyard Transfer Arm
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Stockyard Unit, wait one
minute and then Power ON the Stockyard
Unit to reset.
b. Place Stockyard in AUTO mode to
initialize.
4. If the problem persists, call your Beckman
Coulter Representative.
4-07 SM07 Failure to move the
Stockyard Transfer Arm
on the third level in the
X-axis direction.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure the Stockyard Transfer Arm
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Stockyard Unit, wait one
minute and then Power ON the Stockyard
Unit to reset.
b. Place Stockyard in AUTO mode to
initialize.
4. If the problem persists, call your Beckman
Coulter Representative.
4-08 SM08 Failure to move the
Stockyard Transfer Arm
on the third level in the
Y-axis direction.
3000 Tube
Stockyard
To p V i e w
C-4/5 1. Ensure the Stockyard Transfer Arm
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Stockyard Unit, wait one
minute and then Power ON the Stockyard
Unit to reset.
b. Place Stockyard in AUTO mode to
initialize.
4. If the problem persists, call your Beckman
Coulter Representative.
7-07 AS07 Stockyard Internal
Transfer Arm gripper on
the first level failed to
pick up sample.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure the sample tube is properly
positioned in gripper fingers.
2. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution