Specifications
A64192AA2 FINAL
8-8
Stockyard Unit
Stockyard Unit Error Recovery
2-55 SL55 Stopper error at the
T-Lane before sorting
position.
Stockyard
Unit T Lane
C-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-56 SL56 Stopper error at the
T-Lane sorting position.
Stockyard
Unit T Lane
C-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-57 SL57 Stockyard Divider Arm
error at the entrance of
H Lane.
Stockyard
Unit T Lane
C-3 1. Ensure the gate is free to extend and retract.
2. Ensure the gate arm is not to slow to extend/
retract. If so, the cylinder air flow controller
may need adjustment.
3. Press PAUSE/RUN to clear the error.
2-58 SL58 Stopper error at the exit
of the H-Lane.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-59 SL59 Stopper error at the exit
of the H-Lane.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-60 SL60 Stopper error at the T-
Lane sample tube
carrier sorting position
#1.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-61 SL61 Stopper error at the T-
Lane sample tube
carrier sorting position
#2.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-62 SL62 Stockyard Divider Arm
error at the T-Lane
sample tube carrier
sorting position #2.
Stockyard
Unit T Lane
C-4 1. Ensure the gate is free to extend and retract.
2. Ensure the gate arm is not to slow to extend/
retract. If so, the cylinder air flow controller
may need adjustment.
3. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution