Specifications
A64192AA2 FINAL
8-7
Stockyard Unit
Stockyard Unit Error Recovery
8
2-41 SL41 Stopper error before
the Stockyard bar-code
reader BR01.
Stockyard
Unit
Conveyor
D-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-43 SL43 Stopper error at the
Stockyard bar-code
reader BR01.
Stockyard
Unit
Conveyor
D-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-44 SL44 Stopper error at the
second stopper before
the Stockyard bar-code
reader BR02.
Stockyard
Unit
Conveyor
D-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-45 SL45 Stopper error at the
stopper before the
Stockyard bar-code
reader BR02.
Stockyard
Unit
Conveyor
D-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-46 SL46 Stopper error at the
Stockyard bar-code
reader BR02.
Stockyard
Unit
Conveyor
D-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-48 SL48 Stopper error at the
Stockyard waiting
empty sample tube
carrier queue.
Stockyard
Unit
Conveyor
E-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-49 SL49 Stopper error at the
Stockyard waiting
empty sample tube
carrier queue.
Stockyard
Unit
Conveyor
E-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-50 SL50 Stockyard Divider Arm
error at the sample
tube carrier waiting
lane sorting position.
Stockyard
Unit
Conveyor
D-4 1. Ensure the gate is free to extend and retract.
2. Ensure the gate arm is not to slow to extend/
retract. If so, the cylinder air flow controller
may need adjustment.
3. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution