Specifications

A64192AA2 FINAL
8-6
Stockyard Unit
Stockyard Unit Error Recovery
2-06 SL06 Rotating cylinder error
on the Stockyard
Loading Elevator.
3000 Tube
Stockyard
To p V i e w
E-4
A
1. Ensure the Stockyard Loading Elevator
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-08 SL08 Rotating cylinder error
on the Stockyard
Retrieving Elevator.
3000 Tube
Stockyard
To p V i e w
E-5
B
1. Ensure the Stockyard Retrieving Elevator
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-10 SL10 Stockyard Loading
Shuttle Cup error on
second level.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure the Stockyard Loading Shuttle
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-11 SL11 Stockyard Retrieving
Shuttle Cup error on
second level.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure the Stockyard Retrieving Shuttle
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-12 SL12 Stockyard Internal
Tr a n s f e r A r m a i r
cylinder error on
second level moving up
or down.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure the Stockyard Internal Transfer Arm
air cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-14 SL14 Stockyard Internal Arm
Pushing cylinder error
on the second level.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure that Stockyard Internal Arm Pushing
cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-15 SL15 Stockyard Loading
Shuttle Cup error on
third level.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure the Stockyard Loading Shuttle
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-16 SL16 Stockyard Retrieving
Shuttle Cup error on
third level.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure the Stockyard Retrieving Shuttle
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-17 SL17 Stockyard Internal
Transfer Arm cylinder
error on third level
moving up or down.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure the Stockyard Internal Transfer Arm
cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-19 SL19 Stockyard Internal Arm
Pushing cylinder error
on the third level.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure that Stockyard Internal Arm Pushing
cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-40 SL40 Stopper error before
the Stockyard bar-code
reader BR01.
Stockyard
Unit
Conveyor
D-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid