Specifications
A64192AA2 FINAL
8-5
Stockyard Unit
Stockyard Unit Error Recovery
8
1-63 SN63 Proximity sensor did not
detect a sample tube
carrier on return lane.
Stockyard
Unit T Lane
D--4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-64 SN64 Sample tube carrier
error coming from H-
Lane incoming lane to
make the turn to the
return lane.
Stockyard
Unit H Lane
B-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-65 SN65 Proximity sensor did not
detect a sample tube
carrier turning on H-
Lane.
Stockyard
Unit H Lane
B-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-67 SN67 Stockyard first level
access door open.
3000 Tube
Stockyard
To p V i e w
C-4/5 1. Close the door.
2. Press PAUSE/RUN to clear the error.
3. If the error persists, call your Beckman
Coulter Representative.
1-68 SN68 Stockyard second level
access door open.
3000 Tube
Stockyard
To p V i e w
C-4/5 1. Close the door.
2. Press PAUSE/RUN to clear the error.
3. If the error purists, call your Beckman Coulter
Representative.
1-69 SN69 Stockyard third level
access door open.
3000 Tube
Stockyard
To p V i e w
C-4/5 1. Close the door.
2. Press PAUSE/RUN to clear the error.
3. If the error purists, call your Beckman Coulter
Representative.
2-01 SL01 Stockyard Loading
Shuttle Cup error on
first level.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure the Stockyard Loading Shuttle
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-02 SL02 Stockyard Retrieving
Shuttle Cup error on
first level.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure the Stockyard Retrieving Shuttle
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-03 SL03 Stockyard Internal
Tr a n s f e r A r m a i r
cylinder error on first
level moving up or
down.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure the Stockyard Internal Transfer Arm
air cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-05 SL05 Stockyard Internal Arm
Pushing cylinder error
on the first level.
3000 Tube
Stockyard
To p V i e w
D-4/5
D
1. Ensure that Stockyard Internal Arm Pushing
cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution