Specifications
A64192AA2 FINAL
8-4
Stockyard Unit
Stockyard Unit Error Recovery
1-51 SN51 Proximity sensor did not
detect the expected
sample tube carrier
after it passed BR02.
Stockyard
Unit
Conveyor
D-5 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-52 SN52 Sample tube carrier
error at the Stockyard
waiting sample tube
carrier queue.
Stockyard
Unit
Conveyor
D/E-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-55 SN55 Sample tube carrier
error before the H-Lane.
Stockyard
Unit T Lane
C-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-57 SN57 Sample tube carrier
error at the T-Lane
sample tube carrier
sorting position #1.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-58 SN58 Proximity sensor did not
detect a sample tube
carrier after it entered
the H-Lane.
Stockyard
Unit T Lane
C-3/4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-59 SN59 Sample tube carrier
error on the exit of the
H-Lane.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-60 SN60 Sample tube carrier
error at the T-Lane
sample tube carrier
sorting position #2.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-62 SN62 Proximity sensor did not
detect a sample tube
carrier arriving
Stockyard waiting
empty sample tube
carrier queue.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution