Specifications
A64192AA2 FINAL
8-2
Stockyard Unit
Stockyard Unit Error Recovery
0-21 BZ21 Stockyard bar-code
reader error at BR02.
Stockyard
Unit
Conveyor
D-4 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
0-22 BZ22 Positioning error at the
Stockyard bar-code
reader BR02.
Stockyard
Unit
Conveyor
D-4 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
0-23 BZ23 Invalid route code at
the Stockyard bar-code
reader BR02. Additional
sample programming
caused a sorting
conflict.
Stockyard
Unit
Conveyor
D-4 1. Remove the sample tube.
2. Press PAUSE/RUN to clear the error.
1-04 SN04 Sample tube not
present in the first level
Stockyard Loading
Shuttle Cup.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure sample tube is correctly loaded into
Stockyard Loading Shuttle Cup.
2. Press PAUSE/RUN to clear the error.
1-05 SN05 Sample tube not
present in the first level
Stockyard Retrieving
Shuttle Cup.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure sample tube is correctly loaded into
Stockyard Loading Shuttle Cup.
2. Press PAUSE/RUN to clear the error.
1-21 SN21 Sample tube not
present in the second
level Stockyard Loading
Shuttle Cup.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure sample tube is correctly loaded into
Stockyard Loading Shuttle Cup.
2. Press PAUSE/RUN to clear the error.
1-22 SN22 Sample tube not
present in the second
level Stockyard
Retrieving Shuttle Cup.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure sample tube is correctly loaded into
Stockyard Loading Shuttle Cup.
2. Press PAUSE/RUN to clear the error.
1-32 SN32 Sample tube not
present in the third
level Stockyard Loading
Shuttle Cup.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure sample tube is correctly loaded into
Stockyard Loading Shuttle Cup.
2. Press PAUSE/RUN to clear the error.
1-33 SN33 Sample tube not
present in the third
level Stockyard
Retrieving Shuttle Cup.
3000 Tube
Stockyard
To p V i e w
E-4/5
C
1. Ensure sample tube is correctly loaded into
Stockyard Loading Shuttle Cup.
2. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution