Specifications
A64192AA2 FINAL
8-1
CHAPTER 8
Stockyard Unit
Stockyard Unit Error Recovery
The Stockyard Unit Error Code Table 8.1 describes error messages generated by the LH 1500
Hematology Automation System. The table provides the system error codes, the related sensor, a
location reference point on the schematic diagram and a brief explanation of the possible problem
and possible solutions.
Tab le 8. 1 Stockyard Unit Error Recovery Codes
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution
0-00 BZ00 Bar-code label did not
read at the Stockyard
bar-code reader BR01.
Stockyard
Unit
Conveyor
D-3 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
0-11 BZ11 Bar-code label did not
read at the Stockyard
bar-code reader BR01.
Stockyard
Unit
Conveyor
D-3 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
0-12 BZ12 Positioning error at the
Stockyard bar-code
reader BR01.
Stockyard
Unit
Conveyor
D-3 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
0-13 BZ13 Invalid routing code at
the Stockyard bar-code
reader BR01. Additional
sample programming
caused a sorting
conflict.
Stockyard
Unit
Conveyor
D-3 1. Remove the sample tube and press ENTER.
2. Press PAUSE/RUN to clear the error.