Specifications
A64192AA2 FINAL
6-19
Connection Unit
Connection Unit Error Recovery
6
2-65 SL65 Stopper error at
Conveyor Loading Arm
point "a".
Connection
Unit
Conveyor
D-4
a
1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-66 SL66 Stopper error at
Conveyor Loading Arm
point "b".
Connection
Unit
Conveyor
D-4
b
1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-67 SL67 Stopper error before
Conveyor Unloading
Arm.
Connection
Unit
Conveyor
C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-68 SL68 Stopper error at
Conveyor Unloading
Arm.
Connection
Unit
Conveyor
C-4 1. Clear any obstruction.
2. if a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-69 SL69 Divider Arm error
diverting the sample
carrier with an off-
loaded sample tube
from the Connection
Unit to the conveyor on
the main lane.
Associated with AS69
(retracted) and AS75
(extended).
Connection
Unit
Conveyor
D-5 1. Ensure the Divider Arm is free to extend and
retract.
2. Ensure the Divider arm is not slow to extend
or retract. If so, the flow of air may have to be
adjusted.
3. Press PAUSE/RUN to clear the error.
4. If the problem persists, call your Beckman
Coulter Representative.
2-70 SL70 Conveyor Loading Arm
did not fully extend /
retract while loading
tubes onto the Elevator
Sample Carrier and
misses activating its
limit sensor. The limit
sensors are AS70 (up)
and AS71 (down).
Connection
Unit
Conveyor
C-3 1. Clear any obstruction.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution