Specifications
A64192AA2 FINAL
6-16
Connection Unit
Connection Unit Error Recovery
2-41 SL41 Sample Tube Loading
Arm cylinder did not
extend / retract
properly and misses
activating the limit
sensor. Cassette may
be misaligned. The limit
sensors are AS41
(down) and AS42 (up).
Connection
Unit Side
View
B-3 1. Press PAUSE/RUN to clear the error
2. If the problem persists,:
a. Ensure proper alignment of cassette
position #12 and SL41 finger.
b. Clear any obstruction if possible.
c. Replace covers.
d. Ensure proper alignment of cassette
position #12 and SL41.
e. Clear obstruction if possible.
f. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
2-42 SL42 Sample Tube Loading
Arm cylinder SL42 did
not fully extend / retract
and was unable to
activate its limit sensor.
The sample tube
gripper may have tried
to insert another
sample tube into a
cassette that already
contains a sample tube.
The limit sensors are
AS43 (up) and AS44
(down).
Connection
Unit Side
View
C-3 1. Open the right-side tabletop safety cover.
2. Ensure cassette at loading position #4 is fully
seated to the right in the cassette shuttle.
3. Ensure alignment of the sample tube in the
Sample Tube Loading Arm gripper fingers and
the cassette position immediately in front of
SN49.
4. Replace covers.
5. Press PAUSE/RUN to clear the error.
6. If the problem persists, call your Beckman
Coulter Representative.
2-44 SL44 Sample Tube Unloading
Arm cylinder SL44 did
not fully extend / retract
and was unable to
activate its limit sensor.
Sample Tube Unloading
Arm gripper error
removing the sample
tubes from cassette to
the Unloading Elevator.
The limit sensors are
AS46 (up) and AS45
(down).
Connection
Unit Side
View
B-2 1. Open the right-side tabletop safety cover.
2. Ensure cassette is aligned at unloading
position #3.
3. Ensure alignment of the sample tube in the
Sample Tube Unloading Arm gripper fingers
and the cassette position immediately in
front of SN50.
4. Replace covers.
5. Press PAUSE/RUN to clear the error.
6. If the problem persists, call your Beckman
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution