Specifications
A64192AA2 FINAL
6-14
Connection Unit
Connection Unit Error Recovery
2-23 SL23 Stopped cassette error
at cassette position #6.
Cassette could be
jammed.
Connection
Unit Top
View
D-2 1. Open the left lower front cover.
2. Ensure cassette is properly positioned at
cassette position #6.
3. Replace covers.
4. Press PAUSE/RUN to clear the error.
If unable to remove trapped cassetteā
a. Perform SHUTDOWN at Line Controller.
b. Power OFF automation line at Inlet Unit.
c. Cut off incoming lab air supply to the line.
(If June-air compressor, turn Red Valve in
back of Compressor Unit.)
d. Manually depress air cylinder (SL23) and
then remove trapped cassette.
e. Reposition the cassette in position #6.
f. Reconnect air supply.
g. Power ON the automation line at Inlet
Unit.
h. Perform STARTUP at Line Controller.
5. lf recovery is not possible, call your Beckman
Coulter Representative.
2-24 SL24 The analyzed cassette
is held at cassette
position #7 in the Left
Side Cassette Shuttle.
The Cassette Unloading
Arm cannot release the
cassette.
Connection
Unit Top
View
D-2 1. Press PAUSE/RUN to clear the error.
If arm cannot release cassette:
a. Open the left lower front cover.
b. Press down on the cassette to ensure
cassette releases from Cassette
Unloading Arm.
c. Replace covers.
d. Press PAUSE/RUN to clear the error.
2. If recovery is not possible, call your Beckman
Coulter Representative.
2-25 SL25 Left Side Cassette
Unload Shuttle
movement is
obstructed.
Connection
Unit Top
View
C/D-2 1. Open the left lower front cover.
2. Clear any obstructions.
3. Replace covers.
4. Press PAUSE/RUN to clear the error.
5. If the problem persists, call your Beckman
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)