Specifications

A64192AA2 FINAL
6-12
Connection Unit
Connection Unit Error Recovery
2-03 SL03 During its operation
Cassette Loading Arm
cylinder SL03 did not
swing up/down and
misses activating the
limit sensor. The two
limit sensors for SL03
are AS06 (down) and
AS05 (up).
Connection
Unit
Loading and
Unloading
Arms
B-7 1. Clear any obstruction.
2. Ensure cassette is properly seated on loading
arm fingers.
3. Press PAUSE/RUN to clear the error.
4. If the problem persists, call your Beckman
Coulter Representative.
2-04 SL04 During its operation
Cassette Unloading
Arm cylinder SL04 did
not fully extend and
misses activating the
limit sensor. The two
limit sensors are AS07
(down) and AS08 (up).
Connection
Unit
Loading and
Unloading
Arms
B/C-1 1. Ensure the cassette loading arm fingers align
with sample tube positions #8 and #12 of the
cassette and the positions are empty.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
2-05 SL05 During its operation
Cassette Unloading
Arm cylinder SL05 did
not fully extend and
misses activating the
limit sensor. The two
limit sensors are AS09
(down) and AS10 (up).
Connection
Unit
Loading and
Unloading
Arms
B-1 1. Ensure the cassette loading arm fingers align
with sample tube positions #8 and #12 of the
cassette and the positions are empty.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
2-06 SL06 During its operation
Cassette Unloading
Arm cylinder SL06 did
not swing up/down and
misses activating the
limit sensor. The two
limit sensors are AS12
(down) and AS11 (up.)
Connection
Unit
Loading and
Unloading
Arms
B-1 1. Clear any obstruction.
2. Ensure cassette is properly seated on loading
arm fingers.
3. Press PAUSE/RUN to clear the error.
4. If the problem persists, call your Beckman
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution