Specifications

A64192AA2 FINAL
6-7
Connection Unit
Connection Unit Error Recovery
6
1-37 SN37 Scenarios:
1. Cassette was not
detected inside the
Load Shuttle
cassette position
#3.
2. Cassette incorrectly
detected inside
Load Shuttle during
Startup.
Connection
Unit Top
View
D-7 1. If error occurs when cassette is pushed into
the Load Shuttle:
a. Open the right-side tabletop safety cover.
b. Slide cassette to the right so the SN37
Orange LED is ON.
c. Replace the right-side tabletop safety
cover.
d. Press PAUSE/RUN to clear the error.
2. If the problem occurs immediately after
Startup:
a. Open the right-side tabletop safety cover.
b. Remove cassette from inside the Load
Shuttle cassette position #3.
c. Replace the right-side tabletop safety
cover.
1-38 SN38 One of two limit sensors
associated with
cylinder SL28 Right
Side Load Cassette
Shuttle, position #3.
Connection
Unit Top
View
D-7 This error occurs during Power On, if the sensor is
defective.
If the problem persists, call your Beckman
Coulter Representative.
1-39 SN39 One of two limit sensors
associated with
cylinder SL28 Right
Side Load Cassette
Shuttle, position #4.
Connection
Unit Top
View
C-7 This error occurs during Power On, if the sensor is
defective.
If the problem persists, call your Beckman
Coulter Representative.
1-40 SN40 Problem associated
with moving the
cassette in and out of
cassette position #2
(Tube pushup and
unload area).
Connection
Unit Top
View
C-6 1. Press PAUSE/RUN to clear the error.
2. If the problem persists:
a. Open the right-side tabletop safety cover.
b. Ensure the cassette has traveled from
cassette position #1 to cassette position
#2.
c. Ensure the Tube Unloading Arm finger is
trying to engage the cassette in order to
move it.
3. If the problem persist,
a. When the cassette has already moved to
position #2 and the Load Shuttle is at the
back position #3, then move the cassette
to position #3.
b. Replace the right-side tabletop safety
cover and press PAUSE/RUN to clear the
error.
4. If the problem continues to persist, call your
Beckman Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution