Specifications

A64192AA2 FINAL
6-5
Connection Unit
Connection Unit Error Recovery
6
1-28 SN28 One of two limit sensors
associated with
cylinder SL25. Unload
Cassette Shuttle at
position #8.
Connection
Unit Top
View
C-1 This error occurs during Power ON, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
2. If the problem persists, call your Beckman
Coulter Representative.
1-29 SN29 One of two limit sensors
associated with
cylinder SL25. Unload
Cassette Shuttle at
position #7.
Connection
Unit Top
View
D-1 This error occurs during Power ON, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
2. If the problem persists, call your Beckman
Coulter Representative.
1-30 SN30 One of two limit sensors
associated with
cylinder SL26 at
position #8 at the Left
Side Unload Cassette
Shuttle.
Connection
Unit Top
View
C-2 This error occurs during Power ON, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
2. If the problem persists, call your Beckman
Coulter Representative.
1-31 SN31 One of two limit sensors
associated with
cylinder SL26 at
position #8 at the Left
Side Unload Cassette
Shuttle.
Connection
Unit Top
View
C-1 This error occurs during Power ON, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
2. If the problem persists, call your Beckman
Coulter Representative.
1-32 SN32 Analyzed cassette is in
cassette position #1.
SN32 should point to
sample tube position
#2 in the cassette.
Connection
Unit Top
View
C-4 This error occurs during Power ON, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
2. If the problem persists, call your Beckman
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution