Specifications
A64192AA2 FINAL
6-2
Connection Unit
Connection Unit Error Recovery
1-03 SN03 Obstruction or
hardware error at the
upper limit sensor on
the Cassette Loading
Arm.
Connection
Unit
Loading and
Unloading
Arms
B-5 1. Clear any obstruction.
2. If there is no obstruction:
a. SHUTDOWN at Line Controller and Power
OFF automation line at Inlet Unit.
b. Power ON the automation line at Inlet
Unit.
c. Perform STARTUP at Line Controller.
3. If the problem persists, call your Beckman
Coulter Representative.
1-04 SN04 Obstruction or
hardware error at the
lower limit sensor on
the Cassette Unloading
Arm.
Connection
Unit
Loading and
Unloading
Arms
D-2 1. Clear any obstruction.
2. If there is no obstruction:
a. SHUTDOWN at Line Controller and Power
OFF automation line at Inlet Unit.
b. Power ON the automation line at Inlet
Unit.
c. Perform Startup at Line Controller.
3. If the problem persists, call your Beckman
Coulter Representative.
1-05 SN05 Obstruction or
hardware error at the
middle limit sensor on
the Cassette Unloading
Arm.
Connection
Unit
Loading and
Unloading
Arms
C-2 1. Clear any obstruction.
2. If there is no obstruction:
a. SHUTDOWN at Line Controller and Power
OFF automation line at Inlet Unit.
b. Power ON the automation line at Inlet
Unit.
c. Perform STARTUP at Line Controller.
3. If the problem persists, call your Beckman
Coulter Representative.
1-06 SN06 Obstruction or
hardware error at the
upper limit sensor on
the Cassette Unloading
Arm.
Connection
Unit
Loading and
Unloading
Arms
B-2 1. Clear any obstruction.
2. If there is no obstruction:
a. SHUTDOWN at Line Controller and Power
OFF automation line at Inlet Unit.
b.
Power ON the automation line at Inlet
Unit.
c. Perform STARTUP at Line Controller.
3. If the problem persists, call your Beckman
Coulter Representative.
1-21 SN21 Sample Table Loading
Arm at the home
position did not
initialize due to an
obstruction or hardware
failure.
Connection
Unit Top
View
D-7 1. Clear any obstruction.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution