Specifications
A64192AA2 FINAL
4-2
Inlet Unit Error Recovery
Inlet Unit Error Recovery Codes
1-09 SN09 Sample tube carrier failed
to arrive in the time
allotted at the Inlet sample
tube carrier loading
position.
Inlet Unit C-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently twist
and push the sample tube carrier in the direction
of least resistance.
3. Press
PAUSE/RUN to clear the error.
1-11 SN11 Proximity sensor did not
detect a sample tube
carrier released from the
Inlet sample tube carrier
loading position.
Inlet Unit B-2 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently twist
and push the sample tube carrier in the direction
of least resistance.
3. Press
PAUSE/RUN to clear the error.
1-12 SN12 Sample tube carrier error
before the Inlet sample
tube carrier loading
position.
Inlet Unit C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently twist
and push the sample tube carrier in the direction
of least resistance.
3. Press
PAUSE/RUN to clear the error.
1-13 SN13 Sample tube carrier error
at the Inlet waiting queue.
Inlet Unit C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently twist
and push the sample tube carrier in the direction
of least resistance.
3. Press
PAUSE/RUN to clear the error.
1-20 SN20 Inlet Transfer Arm error
finding the Y-axis home
position.
Inlet Unit C-5 1. Ensure the Inlet Transfer Arm movement is not
obstructed.
2. Press
PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman Coulter
Representative.
1-21 SN21 Inlet Transfer Arm error
finding the Y-axis near
home position.
Inlet Unit C-5 1. Ensure the Inlet Transfer Arm movement is not
obstructed.
2. Press
PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman Coulter
Representative.
2-01 SL01 Inlet Transfer Arm Z-axis
cylinder did not fully
extend / retract and
misses activating the limit
sensor. The limit sensors
are AS01 (up) and AS02.
Inlet Unit Transfer
Arm
1. Clear any obstruction.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman Coulter
Representative.
2-03 SL03 Stopper error at the Inlet
sample tube carrier
loading position.
Inlet Unit B/C-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently twist
and push the sample tube carrier in the direction
of least resistance.
3. Press
PAUSE/RUN to clear the error.
2-04 SL04 Stopper error before the
Inlet sample tube carrier
loading position.
Inlet Unit B/C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently twist
and push the sample tube carrier in the direction
of least resistance.
3. Press
PAUSE/RUN to clear the error.
Tab le 4. 1 Inlet Unit Error Codes (Cont inued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid Ref
Solution