Specifications
A64192AA2 FINAL
2-5
Outlet Unit Error Recovery Codes
Outlet Unit Error Recovery Codes
2
3-01 BR01 Failure of the bar-code
reader BR01 to read.
May have lost its
programming.
Outlet Unit B-3 1. Perform SHUTDOWN at Line Controller.
2. Enter the System Setup > Maintenance
menu.
Default Password=password1
3. Verify correct setting of bar-code label and
exit screen.
Upon exiting the screen the bar-code readers
will be reprogrammed.
4. If the problem persists, call your Beckman
Coulter Representative.
4-01 PM01 Failure to move the
Outlet Transfer Arm in
the X-axis direction.
Outlet Unit E-2 1. Ensure the Outlet Transfer Arm movement is
not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Outlet Unit, wait one minute
and then Power ON the Outlet Unit to
reset.
4. If the problem persists, call your Beckman
Coulter Representative.
4-02 PM02 Failure to move the
Outlet Transfer Arm in
the Y-axis direction.
Outlet Unit B-15 1. Ensure the Outlet Transfer Arm movement is
not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Outlet Unit, wait one minute
and then Power ON the Outlet Unit to
reset.
4. If the problem persists, call your Beckman
Coulter Representative.
a. See Abbreviations for a list of sensor codes.
Tab le 2. 1 Outlet Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution