Specifications

A64192AA2 FINAL
2-4
Outlet Unit Error Recovery Codes
Outlet Unit Error Recovery Codes
2-07 SL07 Stopper error at Outlet
waiting queue lane.
Outlet Unit C-3/4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-08 SL08 Outlet Divider Arm
error at the exit of the
waiting queue.
Associated with AS08
(retracted) and AS09
(extended).
Outlet Unit C-3 1. Ensure the gate arm is free to extend and
retract.
2. Verify the Divider arm is not slow to open and
close. If so, the cylinder air flow controller
may need adjustment.
3. Press PAUSE/RUN to clear the error.
2-09 SL08 #2 Outlet Divider Arm
error at the main lane.
Associated with AS15
(retracted) and AS16
(extended).
Outlet Unit B-3 1. Ensure the gate arm is free to extend and
retract.
2. Verify the Divider arm is not slow to open and
close. If so, the cylinder air flow controller
may need adjustment.
3. Press PAUSE/RUN to clear the error.
2-10 SL10 Stopper error at Outlet
return lane.
Outlet Unit B-5 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-11 SL11 Stopper error at Outlet
return lane.
Outlet Unit B-5 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-12 SL12 Outlet Divider Arm
error at the beginning
of main lane.
Associated with AS12
(retracted) and AS13
(extended).
Outlet Unit B-5 1. Ensure the gate arm is free to extend and
retract.
2. Verify the Divider arm is not slow to open and
close. If so, the cylinder air flow controller
may need adjustment.
3. Press PAUSE/RUN to clear the error.
2-13 SL13 Stopper error before
the Outlet bar-code
reader.
Outlet Unit B/C-3
a
1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
Tab le 2. 1 Outlet Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution